1996 CHRYSLER VOYAGER ECU

[x] Cancel search: ECU

Page 1236 of 1938

CHRYSLER VOYAGER 1996  Service Manual (3) Zero the dial indicator on the cylinder block
mating surface.
(4) Setup the dial indicator on the piston crown
(above the center of the piston pin) 5mm (1/8 in.)
from the edge of the piston and no

Page 1242 of 1938

CHRYSLER VOYAGER 1996  Service Manual (13) Remove pistons and connecting rods from
block.
(14) Remove vibration damper. Refer to procedure
in this section.
(15) Remove front cover. Refer to procedure in this
section.
(16) Remove oil pump

Page 1247 of 1938

CHRYSLER VOYAGER 1996  Service Manual covered during this operation. Mark piston with
matching cylinder number.
(6) Pistons and connecting rods must be removed
from top of cylinder block. Rotate crankshaft so that
each connecting rod is c

Page 1249 of 1938

CHRYSLER VOYAGER 1996  Service Manual assembly, determined by the numbers stamped on
the crown of individual pistons. Engine cylinders are
numbered starting from gear train end of the engine.
Face chamber recess side of piston towards
cam

Page 1255 of 1938

CHRYSLER VOYAGER 1996  Service Manual (3.6196-3.6200 in.). Maximum wear limit .05mm
(.0019 in.).
(2) Check piston pin bores in piston for roundness.
Make 3 checks at 120É intervals. Maximum out of
roundness .05mm (.0019in.).
(3) The pist

Page 1264 of 1938

CHRYSLER VOYAGER 1996  Service Manual a catalytic converter. Failure of the catalytic con-
verter can occur due to temperature increases caused
by unburned fuel igniting when passing through the
converter.
The use of the catalysts also in

Page 1306 of 1938

CHRYSLER VOYAGER 1996  Service Manual (9) Momentarily touch the other end of this
jumper wire to the negative terminal of the battery
for no more than 4 seconds.
(10) Place a rag or towel below the fuel line at the
quick connect to the ra

Page 1343 of 1938

CHRYSLER VOYAGER 1996  Service Manual (8) Verify the electrical connector is attached to
the Proportional purge solenoid (Fig. 50) and not
damaged.
(9) Verify the vacuum connection at the Propor-
tional purge solenoid is secure and not le