2001 INFINITI QX4 engine

[x] Cancel search: engine

Page 1915 of 2395

INFINITI QX4 2001  Factory Service Manual PERFORMANCE CHARTNBHA0031Test ConditionNBHA0031S01Testing must be performed as follows:
Vehicle location Indoors or in the shade (in a well-ventilated place)
Doors Closed
Door window Open
Hood Open
TE

Page 1916 of 2395

INFINITI QX4 2001  Factory Service Manual Both High- and Low-pressure Sides are Too High.NBHA0032S01
Gauge indication Refrigerant cycle Probable cause Corrective action
Both high- and low-pressure sides
are too high.
AC359A
IPressure is reduc

Page 1928 of 2395

INFINITI QX4 2001  Factory Service Manual With Navigation SystemNBHA0103S0102
RHA887H
*1: HA-72 *2: HA-66 *3: HA-45
RHA050G
Ambient Sensor Circuit
COMPONENT DESCRIPTIONNBHA0039The ambient sensor is attached in front of the drivers side con-

Page 1945 of 2395

INFINITI QX4 2001  Factory Service Manual Maintenance of Lubricant Quantity in
Compressor
NBHA0071The lubricant in the compressor circulates through the system with
the refrigerant. Add lubricant to compressor when replacing any
component or

Page 1954 of 2395

INFINITI QX4 2001  Factory Service Manual CHECKING FOR REFRIGERANT LEAKS=NBHA0078IPerform a visual inspection of all refrigeration parts, fittings,
hoses and components for signs of A/C lubricant leakage,
damage and corrosion. A/C lubricant l

Page 1955 of 2395

INFINITI QX4 2001  Factory Service Manual 5. When the A/C operating (compressor running), inject one
bottle (1/4 ounce /7.4 cc) of fluorescent dye through the low-
pressure service valve using dye injector tool J-41459 (refer to
the manufactu

Page 1956 of 2395

INFINITI QX4 2001  Factory Service Manual SHA708EA
3. Move probe along component approximately 25 to 50 mm (1
to 2 in)/sec.
Checking ProcedureNBHA0115S02To prevent inaccurate or false readings, make sure there is no
refrigerant vapor, shop ch

Page 1957 of 2395

INFINITI QX4 2001  Factory Service Manual 5. If a leak detector detects a leak, verify at least once by blow-
ing compressed air into area of suspected leak, then repeat
check as outlined above.
6. Do not stop when one leak is found. Continue